- RIUKAI TECHNOLOGY has 25+ years of experience in the field of environmental testing equipment testing to meet the needs of our customers.
2-Zone Thermal Shock Environmental Chambers
2-Zone Thermal Shock Environmental Chambers
Model: RK-TS2-216
Main Parameters:
Name:Two-Zone Thermal Shock Environmental Chamber
Brand:RIUKAI
High Temperature Tank: +60℃ to +200℃
Low Temperature Tank: A (-55℃ to -10℃); B (-65℃ to -10℃); C (-85℃ to -10℃)
Temperature Fluctuation: ±0.5℃
Heating Time: Ambient temperature to +200℃ ≤ 15 min
Our 2-Zone Thermal Shock Environmental Chamber is a state-of-the-art testing solution, ideal for industries requiring rigorous thermal cycling and shock tests for components, materials, and products. Whether you are in the aerospace, automotive, electronics, or manufacturing sectors, this chamber ensures that your products withstand extreme temperature fluctuations and are tested to meet industry standards for durability and reliability.
Key Features & Specifications:
-
Capacity: 216 L (Large enough for a wide range of testing needs)
-
Inner Box Dimensions: 600 mm x 600 mm x 600 mm (Compact design for flexible testing setups)
-
Impact Temperature Range:
-
High: +60℃ to +150℃
-
Low: Variable -10℃ to -40℃, -10℃ to -55℃, or -10℃ to -65℃ depending on the test requirements
-
-
Heating Time: Ambient to +200℃ in ≤ 15 minutes, providing fast response for efficient testing.
-
Cooling Time:
-
A (+20℃ to -55℃): ≤ 70 minutes
-
B (+20℃ to -65℃): ≤ 80 minutes
-
C (+20℃ to -75℃): ≤ 90 minutes (Ensures precise control of temperature variations)
-
-
Temperature Deviation: ≤ ±2.0℃ (Measured per GB standard, ensuring highly accurate results)
-
Cooling Method: Air-cooled or Water-cooled, offering flexibility for various environments and cooling requirements.
-
Safety Devices: Includes overtemperature protection, voltage protection, compressor safety features, and more, ensuring secure and safe operation during thermal testing.
Applications:
-
Material Testing: Simulate the thermal stresses that materials undergo in real-world conditions, from aerospace components to automotive parts.
-
Electronics Testing: Ensure that electronic components perform reliably even in fluctuating temperatures.
-
Aerospace Testing: Subject your products to the extreme conditions of rapid temperature changes experienced during flight and re-entry.
Advanced Control Features:
-
Pneumatic Switching & Moving Basket: Provides smooth operation for precise temperature cycling.
-
Temperature Fluctuation: ±0.5℃ to simulate the most extreme temperature variations.
-
Impact Recovery Time: ≤ 5 minutes, ensuring quick transition and minimizing downtime.
| Parameter | Specification |
|---|---|
| Model | RK-TS2-216 |
| Capacity | 216 L |
| Inner Box Dimensions (W×H×D) | 600 mm × 600 mm × 600 mm |
| Outer Box Dimensions (W×H×D) | 1600 mm × 2200 mm × 2200 mm |
| Impact Temp. Range | High: +60℃ to +150℃ Low: L (-10℃ to -40℃); F (-10℃ to -55℃); X (-10℃ to -65℃) |
| Tank Temp. Range | High Temp. Tank: +60℃ to +200℃ Low Temp. Tank: A (-55℃ to -10℃); B (-65℃ to -10℃); C (-85℃ to -10℃) |
| Temperature Deviation | ≤ ±2.0℃ (Measured per GB std. 1/10 from wall, ≥ 50mm) |
| Temperature Fluctuation | ±0.5℃ |
| Heating Time | Ambient to +200℃ ≤ 15 min |
| Cooling Time | A (+20℃ to -55℃) ≤ 70 min B (+20℃ to -65℃) ≤ 80 min C (+20℃ to -75℃) ≤ 90 min |
| Test Method | Pneumatic switching, moving basket |
| Impact Recovery Time | ≤ 5 min |
| Basket Movement Time | Within 10 seconds |
| Shell Material | Steel plate (baked paint) or 304 Stainless Steel |
| Liner Material | #304 Stainless Steel Plate |
| Insulation Material | Glass wool or Foamed Polyurethane |
| Heater | Electric heating wire heater |
| Refrigeration Mode | Air-cooled or Water-cooled |
| Compressor | Imported brand |
| Expansion Mechanism | Electronic expansion valve, etc. |
| Refrigerant | High temp side: R404A Low temp side: R508A or R23 |
| Cooler | Built-in closed water tower, external water tower, or built-in condenser |
| Blower | Centrifugal fan |
| Damper Drive | Cylinder |
| Safety Devices | Overtemperature, over/under-voltage & reverse phase protection, compressor overheat/high-pressure/overcurrent protection, buzzer, non-fuse switch, circuit breaker, main power switch, temperature limiter. |
| Power Supply | 220V AC 3Φ 50Hz or 380V AC 3Φ 50Hz |
Notes:
- Performance data is based on testing at 20℃ ambient and 25℃ cooling water temperature with no load.
- Complies with IEC 60068-3-6:2001 / GB/T 2424.6-2006 / JTM K09:2009. Measurement per GB standard.
- Specifications are for reference; actual product specs are subject to the final quotation.